Pressure volume charge testing, high speed operation, compact construction, and mass
flow test technology instrumentation.
A test chamber and a reservoir, approximately equal in volume to the free volume in the test chamber with a part loaded, are pressurized simultaneously using a pre-determined charge volume. The test chamber and the reservoir have a common connection between them. If the part has a leak, air will flow from the chamber into the sealed part and cause the pressure in the chamber to increase. Air will then flow from the reservoir into the test chamber. The flow of air from the reservoir into the test chamber is
directly proportional to the leak in the part and can be accurately measured as the air passes over a transducer.
• Leak range: 0 - 4 sccm.
• Pressure range: 0 - 100 psig (0 - 60 psig, 0 - 150 psig)
• Transducer repeatability: 0.05% F.S.D.
• Transducer sensitivity: 0.05% F.S.D.
• A/D Conversion: 14 bit, 1,000 samples/sec.
• Timer increments: 0.01 sec.
• NIST traceable calibration to less than 5% R&R.
|