Calibration
InterTech CM-15/CM-25 CalMaster
A user friendly transfer standard for the fast and accurate calibration of production leak detectors. NIST traceable. Can be used to calibrate pressure, flow, or vacuum type leak detectors.

The InterTech CM-15 & CM-25 CalMasters are used to induce airflow leaks in calibrating leak detectors. The instruments use a proprietary mass flow sensor to measure the flow of air from the leak detector's calibration port or leak test fixture.


• In calibration mode, the calibration port is connected to the inlet (PRESSURE) port of the CalMaster.
• Air flows through a precision needle valve, across the mass flow sensor and out to atmosphere.
• This flow is measured and displayed on the DRO (digital read out).
• The leak rate is adjusted by varying the setting of the needle valve.
• Calibration of vacuum type leak detectors is similar, except the calibration port is connected to the CalMaster outlet (VACUUM) port, and the inlet (PRESSURE) port is left open to atmosphere.
• Calibration of downstream (bell jar) leak detectors - a supply pressure must be connected to the inlet(PRESSURE) port, while the outlet (VACUUM) port is connected to the calibration port on the leak fixture. This will simulate a leak flow in to the bell jar.

Calibration Accuracy: Traceable to a 0.2% primary standard

Repeatability: 0.05% of Full Scale
Ambient Temperature: 40° F (4° C) to 105° F (40° C)

Proof Pressure Rating: PRES connection: 250 PSIG (17 bar), VAC connection: 5 psig (0.35 bar)

CM-15 CalMaster
Flows under 0.50 sccm, up to 10 psig maximum operating pressure.
Flows between 0.50 - 2.00 sccm, up to 50 psig maximum operating pressure.
Flows between 2.00 - 19.99 sccm, up to 90 psig maximum operating pressure.

CM-25 Micro CalMaster
Flows under 0.500 sccm, up to 10 psig maximum operating pressure.
Flows between 0.500 - 1.999 sccm, up to 50 psig maximum operating pressure.

Available Downloads  
CM-15/25 Spec

 
InterTech Development Company 7401 N. Linder Avenue, Skokie, IL 60077, USA T. 847.679.3377
InterTech Europe 19, rue Clement Michut F-69100 Villeurbanne,France (33)0437.434.116
Shanghai Technical Center 1F, Bldg #36, 458 North Fu Te Road Shanghai Waigaoqiao F.T.Z. 200131 P.R. China, (8621)58682809

 
   
19 rue Clément Michut
69100 VILLEURBANNE
Tel : +33 (0)4 37 43 41 16
Fax : +33 (0)4 37 43 41 17
intertechinfo@intertecheurope.com
  InterTech Development Co
7401 N. Linder Avenue
Skokie, 60077 Illinois USA
TEL 847 -679-3377
FAX 847-679-3391
Service : info@intertechdeveloppement.com
Sales : sales@intertechdeveloppement.com
 